TECHNICAL SPECIFICATION |
|
Ventilation mode and functions S/T T S CPAP AUTO AVAPS Comfort function |
Oscillogram Airway pressure-time waveform graph Flow-time waveform |
Parameter AVAPS: ON/OFF ON/OFF TIDAL VOLUME OFF, 150 ML ~ 1500 mL STEP: 10 mL RESPIRATION RATE 3 bpm ~ 40 bpm Step: 1 bpm INHALATION TIME 0. 3 S ~ 3.0 S step: 0.1 S IPAP S9530: 4 cmH2O ~ 30 cmH2O step: 0.5 cmH2O S9525: 4 cmH2O ~ 25 cmH2O step : 0.5 cmH2O IPAP MAX S9530: 4 cmH2O ~ 30 cmH2O step: 0.5 cmH2O S9525 : 4 cmH2O ~ 25 cmH2O step: 0.5 cmH2O IPAP MIN S9530: 4 cmH2O ~ 29 cmH2O step: 0.5 cmH2O S9525: 4 cmH2O ~ 24.5 cmH2O step: 0.5 cmH2O EPAP S9530: 4 cmH2O ~ 25 cmH2O step: 0.5 cmH2O S9525: 4 cmH2O ~ 20 cmH2O step: 0.5 cmH2O CPAP 4 cmH2O ~ 20 cmH2O Step: 0. 5 cmH2O PRESSURE DELAY RISE TIME 0 ~ 60 min STEP: 1 min PRESSURE RISE TIME STEP: 0.1 S AUTOMATIC ADJUSTMENT, 0.1S ~ 0.6S STEP: 0.1S INSPIRATORY SENSITIVITY AUTOMATIC ADJUSTMENT, 1, 2, 3 , STEP: 1 EXHALATION SENSITIVITY AUTOMATIC ADJUSTMENT, 1, 2, 3 , STEP: 1 COMFORT LEVEL OFF, 1, 2, 3 , STEP: 1 AUTOMATIC FUNCTION ON/OFF Humidifier 0 ~ 5 levels, STEP: 1 |
Alarm and protection Suffocation alarm The suffocation time lasts for 10 seconds and there is no exhalation Pipe disconnection alarm Disconnect the breathing circuit when the flow rate is greater than 60 L/min and lasts for 16 s (±1 s) Low ventilation alarm ~ 30 L/min Range: Off, 1 L/min~30 L/min High respiratory rate alarmrespiratory Range: Off, 4 bpm ~ 80 bpm, and greater than the low respiratory rate alarm setting value Low respiratory rate alarm Range: Off, 4 bpm ~ 40 bpm, and less than the high respiratory rate alarm setting value Airway high pressure alarm T, S, S/T: S9530: 5 cmH2O ~ 30cmH2O S9525: 5 cmH2O ~ 25 cmH2O CPAP: 5 cmH2O ~ 21 cmH2O Airway low pressure alarm T, S, S/T: S9530: 3 cmH2O ~ 29 cmH2O S9525: 3 cmH2O ~ 24 cmH2O CPAP: 3 cmH2O ~ 19cmH2O Low tidal volume alarm When the AVAPS function is turned on Equipment high temperature alarm When the fan temperature is greater than 90 ºC Backup power undervoltage alarm When the battery voltage is lower than 17 V±0.3 V |
Monitor Parameter Freq 0 ~ 80 Bpm Resolution: 1 Bpm V T 0 ~ 2 000 Ml Resolution : 1 Ml MV 0 ~ 40 L/ Min Resolution: 0.1l/ Min When ≥ 10 L/ Min 0.01l/ Min When < 10 L/ Min Tim 0.0 ~ 3.0 S Resolution: 0.1s Breathing Pressure 0 cmH2O ~ 40 cm H2O Step : 0.1 cm H2O |
Working condition Power supply 100 V - 240 V~, 50 HZ/60 HZ, 1.8A-1A (input) DC power supply 24 V (-15% ~ +25%), 3.33A (output) Internal power supply 18.5V, 2850mAh. Ambient temperature range +5ºC~+40ºC Relative humidity range 10~95% Atmospheric pressure range 700kPa~1060kPa |
Certificates
CE,ISO,Free sales certificate,Registeration certificate for medical device,Manufacturing license,qualified industrial registerations & ect .Packing
Transportation
Send your inquiry directly to us